Improved spherical mirror multipass-cell-based interband cascade laser spectrometer for detecting ambient formaldehyde at parts per trillion by volume levels.
AI Summary
Key Findings
Abstract
We report the development of an improved spherical mirror multipass-cell-based interband cascade laser (ICL) spectrometer for ambient formaldehyde (HCHO) detection. The multipass cell consists of two easily manufactured spherical mirrors that are low cost, and have a simple structure, large mirror area utilization, and dense spot pattern. Optical interference caused by the multipath cell was largely reduced, resulting in good sensitivity. Using wavelength modulation spectroscopy (WMS), a detection precision (${1} \sigma $1σ) of 51 pptv in 10 s was achieved with an absorption pathlength of 96 m, which compared favorably with the performance of other state-of-the-art instruments. The precision can be further improved by using a long absorption pathlength configuration and by removing fringe-like optical noise caused by the collimation lens. Ambient application of the developed spectrometer was demonstrated.
External References
- PubMed ID:
- 31873651
- DOI:
- 10.1364/AO.58.008743
Comments
Sign in or create a free account to join the conversation.
Sign in to commentBe the first to comment.