Iodine-125-digoxin radioimmunoassay: comparison of commercial kits.
AI Summary
Key Findings
Abstract
Iodine-125-digoxin radioimmunoassay kits available from Abbott Diagnostics (AD), Dade Division (D), Schwarz/Mann (SM), and Clinical Assays (CA) were evaluated with respect to assay quality. The kit accuracies did not differ significantly at 2.0 ng/ml and the interassay coefficients of variation ranged from 9% (AD) to 21.4% (CA) The accuracy for all kits above 4 ng/ml is questionable, and since serum-dilution values correlated well with undiluted serum values, the dilution method of dose quantitation is preferable for levels above 4 ng/ml. Although all the kits were adequate, for evaluating digoxin at the 2ng/ml level, the Abott kit seems to be of slightly better quality.
External References
- PubMed ID:
- 956901
Comments
Sign in or create a free account to join the conversation.
Sign in to commentBe the first to comment.